Title
The relationship of audiometric thresholds elicited with pulsed, warbled, and pulsed-warbled tones in adults with normal hearing
Abstract
Objective: American Speech-Language-Hearing Association (ASHA) guidelines allow pulsed and warbled tones in measuring audiometric thresholds and include test frequencies of 3000 and 6000 Hz. However, no research has examined the relationship between thresholds obtained with these stimuli at these frequencies. This study investigated the relationship between thresholds obtained with pulsed, warbled, and pulsed-warbled tones.
Design: Thresholds from 25 listeners were obtained using pulsed, warbled, and pulsed-warbled tones at test frequencies recommended by ASHA. RESULTS: Thresholds elicited with pulsed, warbled, and pulsed-warbled tones did not significantly differ.
Conclusions: Findings support using pulsed, warbled, and pulsed-warbled tones for threshold measurements at the frequencies recommended by ASHA.
Department(s)
Communication Sciences and Disorders
Document Type
Article
DOI
https://doi.org/10.1097/AUD.0b013e3181a16366
Publication Date
8-1-2009
Recommended Citation
Franklin, Clifford, Kasey Johnson, Laura Smith-Olinde, and Nannette Nicholson. "The relationship of audiometric thresholds elicited with pulsed, warbled, and pulsed-warbled tones in adults with normal hearing." Ear and hearing 30, no. 4 (2009): 485-487.
Journal Title
Ear and Hearing