The comprehensive validation of a stent coating durability tester
Abstract
This study addresses the need to develop an instrument and supporting protocol to evaluate the tendency for a bare or drug coated stent to shed particles into the vasculature of a patient. The resultant experiments demonstrated that two types of instruments each properly detected, sized and counted better than 90% of the injected calibration particles within four minutes.
Department(s)
Physics, Astronomy, and Materials Science
Document Type
Article
Publication Date
1-1-2009
Recommended Citation
Rajesh, Ramesh, James C. Conti, and Elaine R. Strope. "The comprehensive validation of a stent coating durability tester." Biomedical Sciences Instrumentation 45 (2009): 197-208.
Journal Title
Biomedical Sciences Instrumentation